Functional Diagnostics
ri-thermo® Clibo Temperature Chamber
Riester · 12468
Article No. AK-FD-457
The ri-thermo® Clibo temperature chamber has been specially developed for the regular verification of the measurement accuracy of Riester thermometers and helps you to ensure a consistently high level of patient safety in daily work in a medical practice. Especially for users of the tymPRO/tymPRO+ and sensioPRO/sensioPRO+ models, the temperature chamber enables a quick, standardised and evidence-based accuracy check under reproducible conditions. Thanks to the controlled temperature environment and excellent temperature stability (±0.03 °C at 37 °C), you benefit from reliable and repeatable results, which not only supports medical decision-making but also simplifies compliance with legal and billing-relevant requirements. The target temperatures can be freely selected within the range of 36.0 °C to 42.0 °C and can be displayed in both °C and °F, allowing you to carry out tests exactly according to your individual requirements. Operation is via an intuitive LCD display; integration into the workflow is straightforward and requires minimal time. Another advantage: The temperature chamber is housed in a robust, lockable case, thus meeting the highest safety standards during transport
Specifications
- Brand
- Riester
- Model
- 12468
- Product type
- Thermometers
- Accuracy
- ±0.1 °C / ±0.1 °F
- Temperature stability
- ±0.03 °C at 37 °C
- Measurement unit
- °C (standard) or °F, switchable
- Display
- LCD
- Heating stabilisation time
- 15 to 30 minutes
- Power supply
- 100–240 V (50/60 Hz) / 3 A
- Dimensions (L x W x H)
- 300 x 350 x 160 mm
- Weight
- 4.3 kg
- Case
- suitable for transport, robust, with three-digit combination lock
- Operating conditions
- 16 °C to 32 °C, 30–85 % relative humidity (non-condensing)
- Storage/transport conditions
- –20 °C to 55 °C, 30–85 % relative humidity (non-condensing)
- Nominal emissivity
- 0.98 (suitable for infrared thermometry)
- Certification
- CE marking
- Maintenance
- Calibration check recommended every 2 years