For the provision of a similar maternal uterus environment for the neonate to provide stable and controlled comfortable thermal environment and servo-controlled humidity and oxygen level. It is ideal for use in Infant temperature recovery, resuscitation and other occasions.

Equipment standard by AKLION.

Specifications

Incubator Mode(Close Mode)
Mode of control
Air mode and Baby mode controlled by / micro-computer
Air temperature control range
20°C-37°C(override mode 37°C-39°C)
Baby temperature control range
34°C-37°C(override mode 37°C-38°C)
Accuracy of skin temperature sensor
within ± 0.2°C
The temperature fluctuation between incubator / temperature and average incubator temperature (under stable / temperature condition)
≤0.5°C
Temperature uniformity
≤0.8°C(mattress is on the horizontal position) / ≤1.0°C(mattress is on the maximum tilt position)
Inclination of the bassinet
± 12° (stepless adjustable)
Noise inside hood (stable working condition)
≤50dB(A)
Relative Humidity Level: / Humidity display range / Humidity control range / Accuracy of humidity display / Accuracy of humidity control
0%RH~100%RH / 0%RH~90%RH / ± 5%RH / ± 5%R
Oxygen Level: / Indication range of oxygen concentration / Control range of oxygen concentration / Indication accuracy of oxygen concentration / Control accuracy of oxygen concentration
0%~100% / 20%~60% / ± 2% (The set value is below 25%) / ± 3% (The set value is above 25%) / ±4% 02
Infant Radiant Warmer Mode(Open Mode)
Control Mode
Pre-warm mode, manual mode and baby mode / are controlled by micro-computer
Baby mode temperature control range
34°C-38°C
Control accuracy ( The fluctuation range between Skin temperature and Set temperature)
≤0.5°C
Temperature uniformity of mattress
≤2°C
Performance Index Under Thepulse Oximetry
SPO2 Display range
1- 100%
SPO2 Display resolution
1%
SPO2 Alarm setting range
Upper limit:50 ~100% / Lower limit: 45-95%
PR Display range
25-240 bpm
PR Display resolution
1bpm
PR Alarm Setting range
Upper limit: 80 – 240 bpm / Lower limit: 35-180 bpm
PI Display range
0.02-20.0%
PI Display Resolution
0.01%
Sensitivity setting
Normal,APOD, Max, / Factory default setting: Normal
FastSAT
off, on, Factory default setting: off
Mean time
2S -4S, 4S-6S, 8S,10S, 12S, 14S, 16S / Factory default setting: 8S
FailureAlarm
Power Failure Alarm, Fan Motor Alarm, Sensor, FailureAlarm, Deviation Alarm, CanopyFailure Alarm, OverTemperature Alarm, Water tank placed false alarm, WaterShortage Alarm, SettingAlarnr Checking Alarm, System _ / Alarm, etc.
Transportation
Storage Temperature range: -20°C~+55°C / Relative Humidity range: ≤93% / Atmospheric pressure: 500hPa – 1060hPa
Environment Requirement
Operating range: 20°C-30°C(Set temperature / should be 3°C higher than the environment temperature) / Relative Humidity range: 30%~75% / Ambient air velocity rate: ≤0.3m/s
Product Packaging
Package: The whole unit is packed into one carton / Size: 1140mm x 660mm x 2315mm / Gross Weight: 160Kg
Power supply / Power Requirement / Auxiliary Mains Power Output / Current
AC220V-230V/50Hz / 1300VA / AC220-230V/50Hz, Max / 1A