Anesthesia Workstation Practice 3800 Advanced

Anaesthesiology

Anesthesia Workstation Practice 3800 Advanced

Dixion

Article No. AK-ANA-004

The Anesthesia Workstation Practice 3800 Advanced is part of the Dixion medical equipment range available through AKLION. Contact our team for full technical specifications, configuration options and availability for your facility.

Equipment standard by AKLION.

Specifications

Dimensions (H x W x D) / Trolley version (with breathing circuit)
1420mmx770mmx760mm
Weight: / Trolley (without vaporizer and backup cylinder) Top shelf load
135kg 25kg
Power input
100~240 VAC, 50/60 Hz
Batteries and Operation time with fully charged
DC24V, 4.0AH, Min.120min
Operation temperature
10~+40ºC
Operation humidity
≤95%(non-condensing)
Gas supply
O2, N2O, AIR 280kPa – 600kPa
Cylinder yokes Optional
O2, N2O, AIR
Fresh gas flow indicator
Electronically display and virtual fresh gas flow tubes
Range of fresh gas flow indicators
O2, N2O (0~10.0 l/min), AIR (0~12.0 l/min) / O2flush 25~75 l/min
ACGO
Optional
AGSS
Optional
Agent
Halothane, Enflurane, Isoflurane, Sevoflurane
Installation mode
Selectatec® with interlock, optional standby vaporizer parking holder
Filling type
Pour Fill, Key Fill, Quik-Fil
APL Range
Spontaneous breathing (SP) -70cm H2O
CO2bypass
Optional
Ventilator
Electronically controlled, pneumatically driven
Ventilation modes
Manual/spontaneous / IPPV, PCV, PCV-VG, SIMV-VC, SIMV-PC, SIMV-VG, CPAP,BIVENT APRV
Breathing frequency (rate)
2~100bpm
PEEP
OFF, 3~50cm H2O
Inspiration/expiration ratio (Ti:Te)
4:1~1:8
Tidal volume
20~1500ml(10~1500ml by Volume Control and SIMV modes)(5 to 1500 ml by PressureControl Vent Mode)
Inspiration pause
OFF, 5%~60%
Inspiratory time
0.2~5.0 s
Inspiratory pressure
5~70cm H2O
Pressure support level
3~60cm H2O
Pmax
10~100 cm H2O
Trigger level
0.5~15 l/min
Monitoring
Continuous monitoring of inspiratory O2concentration, breathing frequency, tidal volume, minute volume, peak airway pressure, PEEP, mean or plateau pressure, CO2 concentration (optional), anesthetic gas concentration (optional) paramagnetic oxygen sensors (optional)
Compliance
0~100 mL/cm H2O
Control screen
15 inch; TFT color touch screen
Graph Display
Waveform of P-t, F-t, V-t, CO2-t (option), Paw-V Loop, V-Flow Loop,P-F Loop
Alarm
MV high/low limit, FiO2high/low limit, Paw high/low limit, Power failure Breath Rate high limit / Subatmospheric pressure Sustained airway pressure / Apnea alarm, Alarm Silence(120s)